Teledyne Hastings, manufacturer of the original DV-4 and DV-6 thermocouple gauge tubes, produces vacuum measurement equipment with exceptional stability, accuracy and reliability under the most demanding of conditions. These vacuum sensors are calibrated for nitrogen or air, but conversion factors for other gases are available from the factory.
Specification |
Model
HPM 2002
|
Model
HPM-2002 OBE |
|
|
|
View Product
Bulletins for
Ordering Codes |
|
|
Manuals |
|
|
Drawings |
- |
- |
Model Number |
HPM 2002 |
HPM 2002 OBE |
Vacuum Range |
0.0001 - 1000Torr
(0.1 mTorr - 1000 Torr) |
0.0001 - 1000Torr
(0.1 mTorr - 1000 Torr) |
Accuracy |
+/-1.5% of Reading 100-1000 Torr
+/-20% of Reading 0.001 - 100 Torr |
+/-1.5% of Reading 100-1000 Torr
+/-20% of Reading 0.001 - 100 Torr |
Sensor Type
Direct Force |
Piezo resistive - 32 - 1000 Torr
Independent of Gas Composition
Weighted AVG between 8 and 32 Torr |
Piezo resistive - 32 - 1000 Torr
Independent of Gas Composition
Weighted AVG between 8 and 32 Torr |
Sensor Type
Thermal
Conductivity |
Pirani - 0.0001 - 8 Torr
Dependent on Atmosphere
Weighted AVG between 8 and 32 Torr
|
Pirani - 0.0001 - 8 Torr
Dependent on Atmosphere
Weighted AVG between 8 and 32 Torr |
Sensor
Part Numbers |
(sold separately)
HPM-2002S-01 1/8" NPT (STD)
HPM-2002S-02- 1/4" VCR
HPM-2002S-03- 1.33 Conflat
Flange
HPM-2002S-04- 2.75 Conflat Flange
HPM-2002S-05-KF-16
HPM-2002S-06-KF-25
HPM-2002S-07- 1/2" OD Smooth Tube
HPM-2002S-08- 1/2" VCR
|
(included)
1/8" NPT (STD)
1/4" VCR
1.33 Conflat Flange
2.75 Conflat Flange
KF-16
KF-25
1/2" OD Smooth Tube
1/2" VCR |
Display |
3 Digit LED floating point mantissa
Plus 2 digit exponent
|
4 Digit LED floating point |
Units |
User Selectable -Torr, mbar, Pa |
Torr (STD)
mbar
Pa
|
Gas Displayed |
Nitrogen (STD)
Argon
Helium
Water Vapor
Custom
|
Nitrogen (STD)
Argon
Helium
Water Vapor
Custom |
Calibration |
3 point (STD)
NIST with Documentation
NIST with Certificate (no
documentation)
Custom Calibration
|
3 point (STD)
NIST with Documentation
NIST with Certificate (no
documentation)
Custom Calibration |
Calibration Gas |
Nitrogen |
Nitrogen |
Panel Cutout
Size |
1/4" DIN
W-3.75"
H-3.75"
D-6.60"
|
- |
Standard Output |
Analog - 1.0 - 4.5 Volts
Single Channel
Linear per decade
|
Dual Analog 0 - 10 Volts
2 Channel
Linear
|
Optional
Outputs |
RS 232
RS-485
Dual 4-20 mA (with or without Control
Relay (1A)
Dual 0-10 VDC (with or without
Control
Relay (1A)
Dual Control Relays
Device Net® |
RS 232
RS-485 Half Duplex
RS-485 Full Duplex
Dual 4-20 mA
Device Net®
|
Control Points
STD
|
2 TTL Outputs
1 TTL Remote Zero Input
|
2 TTL Outputs
1 TTL Remote Zero Input
|
Operating
Temperature
Range
|
0o- 50o C |
0o- 50o C |
Response Time |
200 milliseconds |
200 milliseconds |
Power |
90 - 260 VAC - 50 - 60 Hz |
13 -30 VDC (power supply available) |
Power Cable
Length
|
6 feet |
Optional Power Supply Cord |
Vacuum Sensor
Cable Length
|
6.2 feet (2 meters)
15.5 feet (5 meters)
(purchased separately)
|
- |
Weight
(approx) |
2.3 lbs (1 kg) |
12 oz (with 1/8" NPT) |