Specification |
Mass Flow Meter & Controller
HFM-D-300
HFC-D-302 |
Mass Flow Meter & Controller
HFM-D-301
HFC-D-303 |
Mass Flow Meter & Controller
HFM-D-305
HFC-D-307
|
| View Product Bulletins for ordering codes |
Find Ordering Codes On Last Page |
Find Ordering Codes On Last Page |
Find Ordering Codes On Last Page |
| Manuals |
HFM-D-300 /
HFC-D-302 |
HFM-D-301 /
HFC-D-303 |
HFM-D-305 /
HFC-D-307 |
| Drawings |
HFM-D-300 /
HFC-D-302 |
HFM-D-301 /
HFC-D-303 |
HFM-D-305 /
HFC-D-307 |
| Full Scale Flow Range |
0-5 sccm to
0-25,000 sccm |
0-25 slm to
0-1000 slm |
0-1000 slm to
0-3,000 slm |
| Accuracy |
+/- (0.5% of reading + 0.2% of Full Scale) |
+/- (0.5% of reading + 0.2% of Full Scale) |
+/- (0.5% of reading + 0.2% of Full Scale)
+/-2% of full scale for 2,500 - 3,000 slm |
| Repeatability |
+/-0.1% of Full Scale |
+/-0.1% of Full Scale |
+/-0.1% of Full Scale |
| Standard Pressure Rating |
500 psig |
500 psig |
500 psig |
| Pressure Coefficient |
+0.01%/psi of reading (0-50 psig N2) typical
|
+0.01% of reading/psi (0-1000 psig N2) typical |
+0.015%/psi(0-500 psig N2) |
| High-Pressure Option |
1000 psig (proof tested to 1500 psig) |
1000 psig |
Not Available |
| Temperature Coefficient of Meter |
Zero <0.2%/o C of Full Scale
Span <0.05%/ o C of Reading |
Zero <0.2%/o C of Full Scale
Span <0.05%/ o C of Reading |
Zero <0.2%/o C of Full Scale
Span <0.05%/ o C of Reading |
| Temperature Coefficient of Controller |
Zero insignificant when auto zero is active
Span <0.05%/ o C of Reading
|
Zero insignificant when auto zero is active
Span <0.05%/ o C of Reading |
Zero insignificant when auto zero is active
Span <0.05%/ o C of Reading |
| Leak Integrity |
< 1 x 10-9 sccs He |
1 x 10-9 sccs He |
1 x 10-9 sccs He |
| Power - Meter |
+/-15VDC at 55 mA |
+/-15VDC at 55 mA |
+/-15VDC at 55 mA |
| Power - Controller |
+/-15 VDC at 150 mA |
+/-15 VDC at 150 mA |
+/-15 VDC at 150 mA |
| Flow Signal |
0-5.00 VDC, 0-10 VDC
0-20 mAmp, 4-20 mAmp |
0-5.00 VDC, 0-10 VDC
0-20 mAmp, 4-20 mAmp |
0-5.00 VDC, 0-10 VDC
0-20 mAmp, 4-20 mAmp |
| Attitude Sensitvity of Zero |
< 0.7% of Full Scale (N2@19.7 psia) |
< 0.7% of Full Scale |
< 0.7% of Full Scale |
| Attitude Sensitvity of Span |
< 0.05% of Full Scale (N2@19.7 psia) |
< 0.05% of reading |
< 0.05% of reading |
| Settling Time of Meter |
</= 0.5 sec (0% to 100% FS) |
</=0.5 sec (0% to 100% FS) |
</=0.5 sec (0% to 100% FS) |
| Settling Time of Controller |
</= 0.75 sec (0% to 100% FS) |
</= 2.0 sec (0% to 100% FS) |
</= 2.0 sec (0% to 100% FS) |
| Wetted Material - Meter |
302 SS, 316 LSS, Ni 200 |
Viton® , 302 SS, 316 LSS, Ni 200 |
302 SS, 316 LSS, NI 200 |
| Wetted Material - Controller |
302 SS, 316 LSS, Ni 200, Kalrez ® , (Valve Seat) |
Viton® , 302 SS, 316 LSS, Ni 200, Teflon ® ,(Valve Seat) Kalrez® |
302 SS, 316 LSS, Ni 200, Kalrez ® (Valve Seat) |
| Connector |
15-pin subminiature D |
15-pin subminiature D |
15-pin subminiature D |
| Digital Communication |
RS 232 (STD)
RS 485 |
RS 232 (STD)
RS 485 |
RS 232 (STD)
RS 485 |
| Pin out diagrams |
H - Pin out |
H - Pin out |
H - Pin out |
| Fittings |
1/4" Swagelok® (STD) Options
1/8" Swagelok®
1/4" VCR® pair
1/4" VCO® pair
6mm Swagelok® pair
No Fittings |
1/2" Swagelok® (STD) Options
1/2" VCR® pair
1/2" VCO® pair
12mm Swagelok® pair
No Fittings |
1" Swagelok® (STD) Options
1" VCR® pair
1" VCO® pair
25mm Swagelok® pair
No Fittings |
| Weight (approx) |
HFM 300 - 1.95 lbs
HFC 302 - 2.45 lbs |
HFM 301 - 3.5 lbs
HFC 303 - 5.3 lbs |
HFM 305 - 8.2 lbs
HFC 307 - 15.3 lbs |